0

Gel Timer -1 Channel Control Unit with Temperature Measurement, incl. software

Model Number: PST-1

1 Channel Control Unit with Temperature Measurement, incl. software as per DIN 16 945 page 1, DIN 16 919, DIN EN ISO 9396 and DIN EN 14022 method 5

The Gelnorm®– Geltimer PST-1 evaluates all factors which influence the gel time of reaction resins such as type and quantity of hardener, accelerator, inhibitor, filler, pigments, as well as temperature and moisture. The instrument is easy to operate and normally no maintenance necessary.

The principle of measurement of Gelnorm®– Geltimer is simple: A stamper made of aluminium or stainless steel performs an up-down cycle in a test tube filled with resin. When the point of gelation is reached, the test tube is pulled up by the stamper. This stops the clock which was started at the beginning of the experiment and the gel time can be read of. With the probe temperature system the exothermic reaction can be measured. The electrical heating device, option heating block GT, can be controlled directly with the internal temperature controller.

Qty :

INDUSTRY: Other

APPLICATION: Gelation Time

Case Dimension: 1

Temperature control for the electrical heating block GT

Actual value (Pt100) Range: 0,0 °C ... 250,0 °C, adjustable and calibrated
Accuracy ± 0,3% of full scale ± 1 digit
Set value Range: 0,0 °C ... 200,0 °C (on request 250,0 °C)
Resolution 0,1 K
Load output power max. - resistive load : maximum 600 W


Temperature measurement for the sample
Sensor Thermocouple type K (Ni- CrNi) , EN 60 584
Connector Mini connector type K thermocouples, EN 60 584
Range 0,0 °C … 400,0 °C
Accuracy ± 1,5°C including cold-junction compensation

Time
Display LCD, 12mm
Start Button with status indicator
Stop optical switch
Range up to 99HH: 59MM: 59SS
Accuracy ± 1 second / 2 hours

Leaflet

Add a Review